Abort test sequence if fails
HV Terminal used (AC-30 not used or not installed) since both fields 15 and 16 are missing
DUT is isolated since both fields 15 and 16 are missing
ACI CONFIGURATION FIELDS
Field # Field Format Value
1
<STRING>
ACI
2
<NR3>
Test time (in seconds)
<EMPTY>
Step is to be user terminated
3
<NR3>
Delay time (in seconds)
4
<NR3>
Minimum Leakage Current Limit (in Arms)
5
<NR3>
Maximum Leakage Current Limit (in Arms)
6
<NR1>
Time period for arc detection (4, 10, 15, 20, 30 or 40)
<EMPTY>
Arc detection disabled
7
<NR1>
Arc detection limit (in mA)
<EMPTY>
Only valid if arc detection disabled in field 6
8
<STRING>
ABORT - Abort on failure
CONT - Continue on failure
Example
–
ADD,ACI,5.0,0.0,0.0,50e-6,,,ABORT
Configures the following (in order)
–
An ACI type test step
5 second test time
0 second delay time
No minimum leakage current
50uA maximum leakage current
No arc detection (so following field is also <EMPTY>)
Abort test sequence if fails
DCI CONFIGURATION FIELDS
Field # Field Format Value
1
<STRING>
DCI
2
<NR3>
Test time (in seconds)
<EMPTY>
Step is to be user terminated
3
<NR3>
Delay time (in seconds)
4
<NR3>
Minimum Leakage Current Limit (in A)
5
<NR3>
Maximum Leakage Current Limit (in A)
6
<NR1>
Time period for arc detection (4, 10, 15, 20, 30 or 40)
<EMPTY>
Arc detection disabled
7
<NR1>
Arc detection limit (in mA)
<EMPTY>
Only valid if arc detection disabled in field 6
8
<STRING>
ABORT - Abort on failure
CONT - Continue on failure