5 First steps
17
Carrying out the configuration
Requirement: The instrument is switched off.
1. Switch the instrument on.
> Switch the instrument on, press
[Hold]
during the display
test and hold down for approx. 2 secs.
- When the configuration menu is locked,
PIN
lights up.
> To unlock, set the last two figures of the serial number with
[
]
or
[ ]
.
If an incorrect PIN is entered, the instrument switches into
measuring mode.
- When the configuration menu is not locked, you can gain
access by setting the limit values.
2. Set the alarm limit values.
-
Alarm
lights up in the display.
> Set lower limit value (
Alarm
):
[ ]
or
[ ]
and confirm with
[Hold]
.
-
Alarm
lights up in the display.
> Set upper limit value
(Alarm )
:
[ ]
or
[ ]
and confirm with
[Hold]
.
3. Carry out / do not carry out calibration.
-
CAL
and
no
or
yes
light up in the display.
> Activate / deactivate calibration / adjustment function (
no
is
the standard setting):
[ ]
or
[ ]
and confirm with
[Hold]
.
- Selection of
yes
: The instrument can be calibrated /
adjusted,
- Selection of
no
: No calibration / adjustment possible
4. Automatically hold readings.
-
Auto-Hold
and
on
or
off
light up in the display.
> Switch
Auto-Hold
on or off:
[ ]
or
[ ]
and confirm with
[Hold]
.
5. Automatically switch the instrument off.
-
Auto-off
and
on
or
off
light up in the display.
> Switch
Auto-off
on or off:
[ ]
or
[ ]
and confirm with
[Hold]
.