6 Using the product
23
3. To hold readings: Briefly press
[Hold]
(< 1 sec).
-
Hold
appears in the display.
- Readings are held.
4. Read the measurement values.
5. To switch to the measuring mode: Briefly press
[Hold]
(< 1 sec).
6.3.
Function test
For a simple function test without adjustment (accuracy
+/- 3 % TPM
2
), we recommend a measurement during the
commissioning of your new instrument in unused deep-frying oil at
150 to 180 °C.
We recommend performing the function test every time after
refilling the deep fryer with fresh oil.
1.
Perform measurement in unused deep-frying oil at 150 to
180 °C, see
Carrying out measurements, page
21
.
2.
Note reading.
3.
Repeat steps 1 and 2 several times.
- The average of the readings is your specific reference value for
subsequent instrument testing.
When changing the type of oil or the oil supplier, the
reference value must be determined anew.
Your specific reference value:
___________________________
2
typical, referred to internal reference, at an ambient temperature of 25 °C