8 Tips and assistance
26
8
Tips and assistance
8.1.
Questions and answers
Items shown on the
display:
Possible causes / solution
lights up and flashing
temperature value
<40 °C
appears in the display
Permissible measuring range
undershot
> Increase oil temperature.
lights up and flashing
temperature value
>
200 °C
appears in the
display
Permissible measuring range
exceeded
> Reduce oil temperature.
Battery symbol
lights
up
Battery charge level low (approx.
2.5 hrs remaining life)
> Change batteries if necessary.
Battery symbol
flashes Batteries empty
(approx. 1.2 hrs remaining life)
> Change batteries.
000
lights up
Sensor not in oil
> Instrument is ready for
measurement. Immerse sensor in
oil.
PIN
lights up
Configuration mode locked.
> Unlock configuration mode.
Err 1
lights up
TPM sensor faulty
> Contact Vito AG or your dealer.
Err 2
lights up
Temperature sensor faulty
> Contact Vito AG or your dealer.
Err 3
lights up
TPM sensor and temperature sensor
faulty
> Contact Vito AG or your dealer.
Err 4
lights up
Other fault
> Contact Vito AG or your dealer.