Vitalograph In2itive User Manual
Cat. No. 07514
Issue 5
Page 9 of 45 DT_0006-6
the top cover. The Vitalograph font and button on the device cap
should be on the same face as the LCD when assembled. Ensure
this is fully pushed in.
5. Attach the remote flowhead adaptor to the flowhead. This is done
by sliding the flowhead into the grooves in the remote flowhead
adaptor. Ensure this is fully pushed in.
6. Open the rubber cover on the left side of the device cap. This will
expose the two pressure ports on the device cap.
7. Attach the remote flowhead to the Vitalograph In2itive device cap
by the dual silicone tubing (Flowhead Connection Tube).
8. A recess is provided on the cradle to hold the remote flowhead
when not in use.
9. The remote flowhead can be disconnected by removing the device
cap from the Vitalograph In2itive, and removing the remote
flowhead adaptor from the flowhead. Then slide the flowhead back
onto the Vitalograph In2itive.
10. It is recommended that an accuracy check is carried out after the
remote flowhead is fitted or removed to verify correct operation and
accuracy.