Vitalograph In2itive User Manual
Cat. No. 07514
Issue 5
Page 21 of 45 DT_0006-6
Note: If the device is outside calibration you will be given the
option to update the calibration. If you select this option you will
be brought through the accuracy check routine again.
When to Check Accuracy
•
In accordance with your own established procedures
•
After annual maintenance checks
•
After cleaning or disassembling spirometer for any reason
•
After adjusting calibration
•
If the flowhead or device has been dropped.
Configuration Options
There are a number of Configuration options available on the
Vitalograph In2itive device. To access these, select the ‘Configuration’
option on the
Main Menu
. The options available are:
Test Preferences
This allows you to configure the test screen to your requirements. The
following options are available:
d) FVC Display: You can select to show F/V (Flow-Volume) or V/T
(Volume-Time) Graph by default in the FVC test screen. Select the
required option from the drop down menu.
e) VC Display: You can select to show the Bar Graph or V/T
(Volume-Time) Graph by default in the VC test screen.
f) Test Acceptance: This allows you to manually accept the tests
performed, or allow the device to determine test acceptability
(automatic).
g) Graph Scale: This allows you to select the default graph scale.
h) Post VC Test: If you have done a VC test in a pre test session and
then go to do a Post test on that session then the VC test screen
will automatically come up. By selecting ‘off’ in the drop down list
you will be brought directly into the FVC test screen.
i) Posture: You can select to no posture selected, standing or sitting.
j) Temperature: By selecting ‘on’ in the drop down list, the device
give the user the option to manually enter the ambient
temperature as they go into the VC or FVC test screens.