Vitalograph In2itive User Manual
Cat. No. 07514
Issue 5
Page 39 of 45 DT_0006-6
Guidance and manufacturer’s declaration – electromagnetic immunity
The Model 2120 is intended for use in the electromagnetic
environment specified below. The customer or the user of the Model
2120 should assure that it is used in such an environment
Immunity test
IEC 60601
Test level
Complian
ce
level
Electromagnetic
environment -
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±
6 kV contact
±
8 kV air
±
6 kV
contact
±
8 kV air
Floors should be
wood, concrete or
ceramic
tile.
If
floors are covered
with
synthetic
material,
the
relative humidity
should be at least
30%.
Electrical
fast
transient/burst
IEC 61000-4-4
±
2kV for power
supply lines
±
1
kV
for
input/output lines
500V
See Warning 2
below.
Surge
IEC 61000-4-5
±
1kV
differential
mode
±
2 kV common
mode
±
1kV
differential
mode
Voltage
dips,
short
interruptions
and
voltage
variations
on
power
supply
input lines
IEC 61000-4-11
<5 % 100V
(>95% dip in
100V) for 0.5 cycle
40 % 100V
(60% dip in 100V)
for 5 cycles
70 % 100V
(30 % dip in 100V)
for 25 cycles
A
A
A
Unit has a battery
installed