Database
This allows you to manage the available memory on the device. The
Management tab tells you how much subject and test session memory
has been used up. As with all memory after a period of time and
repeated use the memory becomes fragmented. Because of this the
total available memory is not being used. In order to correct this, select
the Defragment option. This process may take several minutes to
complete.
The Deletion tab allows the user to delete sessions. By selecting the
relevant box the session will be deleted from the device after being
Printed. (Either a print to an external printer or sent to Vitalograph
Reports).
You are also given the option to ‘Select Test Session(s)’ for deletion, or
‘Delete Subject(s)’ from the device by pressing the relevant button
using the touch panel LCD.
Calibration
The Vitalograph In2itive should never be outside accuracy limits unless
damaged or in a fault condition. In this event, see the fault-finding
guide. In normal use, calibration traceability certification is
recommended as a part of the routine annual service.
Select the Calibration Options menu. You are presented with three
options:
-
Precision Syringe
-
Calibration
Precision Syringe:
b) Select Precision Syringe from the
Calibration
screen.
c) Select the volume of the calibrated syringe you are
using from the drop down list.
d) Enter the syringe reference number using the touch
panel keypad.
e) Press ‘Enter’ to save the new volume entered and
return to the
Calibration
screen. Press ‘C’ to cancel
the changes made and return to the
Calibration
screen.
Calibration
a) Select Calibration from the
Calibration
screen.