Vitalograph In2itive User Manual
Cat. No. 07514
Issue 5
Page 34 of 45 DT_0006-6
routine.
Problem Fault
Symptoms:
•
Rocking cradle
Possible Causes:
(In probable
order)
•
Check for damaged or missing rubber feet.
•
If any of the rubber feet are damaged or
missing replace all rubber feet.
Problem Fault
Symptoms:
•
Reversed or no volume measurements.
Possible Causes:
(In probable
order)
•
Ensure that the flowhead connecting tube is
not pinched or trapped.
Problem Fault
Symptoms:
•
Cannot print to external printer.
•
Corrupt or missing data on printout.
Possible Causes:
(In probable
order)
•
Check that external printer is selected in the
Report Options screen.
•
Check USB cable is connected between
In2itive Cradle and printer.
•
Check
printer
as
per
manufacturers
instructions.
•
Check printer compatibility – contact support.
•
Main PCB failure – contact support.
Problem Fault
Symptoms:
•
Cannot print to PC (Vitalograph Reports
Application).
•
Corrupt or missing data on printout.
Possible Causes:
(In probable
order)
•
Check that Send to PC is selected in the
Configuration screen.
•
Check USB cable is connected between
Vitalograph In2itive and the PC.
•
Check to ensure the Vitalograph Reports
Application is correctly installed.
•
Check to ensure the required software drivers
are installed on the PC.
•
Main PCB failure – contact support.
Problem Fault
Symptoms:
•
Cannot communicate with Spirotrac V