Table of Materials Used & Cleaning/Disinfection Methods
This listing of materials used is given to provide users with information
to allow the assessment of other cleaning and disinfecting procedures
available in the facility on this device.
Part
Material
Clean/
Disinfect
Autoclave
Possible?
Recommended
Disinfectants
Case
Exterior
PC/ABS
Clean
No
White
Flowhead
Tube
Silicone
Rubber
Clean
Viable
Remote
Flowhead
Attachments
PC/ABS,
Silicone
Rubber
Clean
No
Cradle Case
Exterior
PC/ABS
Clean
No
Stylus
PC/ABS
Clean
No
Screen
Electrode
with Anti-
Newton-
Ring
Treatment
Clean
No
Wiping
with
a
70%
isopropyl
alcohol
impregnated cloth provides
a suitable form of cleaning
and low-level disinfection.
This may be preceded by
cleaning with an anti-static
foam cleaner if necessary.
Flowhead
PC/ABS,
Stainless
Steel.
Clean
No
Flowhead
Cone
PC/ABS
Clean
No
Disinfect by immersion in
sodium
dichloroisocyanurate
solution at 1000 ppm
concentration of free
chlorine for 15 minutes.