IFU-08, Rev D | Date: November 2019
Page 31 of 39
e. When pairing is complete, the Blood Pressure section on the VistaPoint application displays the
paired Bluetooth ID. The Model UA-651 screen displays “End” before shutting off. The Model UA-
651 may now be used to upload blood pressure readings to the VistaPoint application.
Figure 41: Pairing Complete on the Model UA-651 and the VistaPoint application
Taking a Spot-Check Reading with the Model UA-651
Use the following procedure to take a spot-check blood pressure reading with a Model UA-651 that has
previously been paired with the VistaPoint application.
Step 1: Launch VistaPoint
If it is not currently running, launch the VistaPoint application. Ensure the VitalPatch biosensor on the
patient is paired with VistaPoint and is currently displaying the patient’s vital signs.
Step 2: Select Take Reading
In the Blood Pressure section, tap the
Take Reading
button. The Blood Pressure section displays a
message indicating that it is ready to take a reading.
Step 3: Apply Cuff and Start Measurement
Apply the Model UA-651 cuff to the patient’s arm and press the START button on the base unit: