TVC Counting Scale Technical Manual (v1.09b-v1.11)
- 20 -
7.2 Normal Calibration
1.
Turn on the scale.
2.
Press and hold
key during the self-test.
Display will show
“
Pn
”
briefly PIN is activated. Press the PIN number to get into
the parameter menu. The default PIN number is “9999”
3.
Press numeric key 9 four times, display will be shown
Pn - - - -
4.
Press
key to confirm, It will enter into the parameter settings and show
“
menu unit lb
” .
a.
If calibrating in
lb
, continue step 6
b.
If calibrating in
kg
, press
key,
kg
will be show on the display
5.
Make sure there are no loads on the platform
PS: Press
key to choose "kg" or "lb" unit, capacity will be show
in kg or lb values.
6.
Press
key to confirm, display will be show “
F0 CAL
”
Press calibration switch to enable settings; the calibration switch is located
below the scale.
7.
Press
key to confirm for calibration, display will be show “
unLoAd”
8.
Remove the goods from the platform and make sure STABLE indication has
been enabled.
(the test weights must be above up to full capacity.)
9.
Press
key to confirm, display will be show
“load”
10.
Calibration weight defaults to scale range and place the calibration weight on
the platform.