AVX-10K Operation
Rev. 002
2022-08-25
22135254
VIAVI
Page
76
of
139
When running this test, the returned values can include:
•
INTERVAL TIME (squitter has been captured)
•
NO SQUITTER (squitter not captured)
•
NO REPLY (no squitter was received)
•
NOT RUN (test has not yet been run)
•
BAD SETUP – This can happen for a number of reasons:
1.
An ADS-B Version Number earlier than DO-260B is being broadcast
2.
The RF input was overloaded at the beginning of the test (usually indicates that
antenna range/height and/or cable loss values are incorrect.
If you choose to save your results, press the Save Data softkey and please refer to Chapter 3 for
managing reports.