AVX-10K Operation
Rev. 002
2022-08-25
22135254
VIAVI
Page
102
of
139
Setting
Definition
RF Port
The VOR Instrument supports either over-the-air antenna
testing or direct connection to the LRU.
RF Power Units
Can be set to dBm or V (Volts in 50 ohms)
Frequency Input Type
PRESET, CHANNEL OR VAR:
PRESET provides a set of three fixed frequencies;
CHANNEL allows the selection of any on-channel frequency;
VAR allows selection of frequency in 1kHz steps
External Attenuation
This parameter allows for compensation (0.0 to 22.0 dB) of an
External Attenuator on the RF I/O Connector. This is used for
extending the power monitor rating and range. If the External
Attenuator is selected, the value is displayed on all Mode
Screens in the EXT ATTN Field under the RF LVL Field.
Morse Code
Up to four characters that are used to transmit Morse Code for
ILS Localizer
The main screen of the VOR application also contains configurable parameters. These configurable
fields have a blue background.
6.2.2. Run VOR Application
Once you enter the VOR Application, the VOR Instrument is running. Changes to the Instrument
Configuration can be made at any time.
Let’s review the configurable fields in the main body of the VOR Application.
Содержание AVX-10K
Страница 1: ...AVX 10K Flight Line Test Set Operations Manual ...
Страница 17: ...AVX 10K Operation Rev 002 2022 08 25 22135254 VIAVI Page 15 of 139 Figure 6 Side View ...
Страница 86: ...AVX 10K Operation Rev 002 2022 08 25 22135254 VIAVI Page 84 of 139 ...
Страница 88: ...AVX 10K Operation Rev 002 2022 08 25 22135254 VIAVI Page 86 of 139 ...
Страница 109: ...AVX 10K Operation Rev 002 2022 08 25 22135254 VIAVI Page 107 of 139 ...