Troubleshooting
6-3
Table 6-1 EX3000 Anesthesia System Problem – Solution Matrix (continued)
No.
Problem
Cause
Solution
5
Needle on system
pressure gauge
(Figure 1-5, 3)
indicates a negative
pressure
a. Excessive
scavenger
system vacuum
Decrease scavenger system vacuum. Recommend
a scavenger interface valve to prevent removal of
patient gases. Also see items 9c and 10c.
b. Inadequate fresh
gas flow
Increase fresh gas flow rates.
6
Patient sleep level
seems too light
a. Vaporizer is
empty
Fill vaporizer with the appropriate anesthetic.
b. Anesthetic
concentration is
set too low
1. Press the anesthetic agent concentration
section (Figure 1-5, 2) on the touch screen.
2. Press
S
or
T
until the desired concentration is
displayed.
c. Leak in patient
breathing circuit
Perform a System Pressure Leak Test as specified
below.
d. Excessive CO
2
build-up
1. Check CO2 absorbent and replace if necessary.
2. Check proper function of inspiration and
expiration valves (Figure 1-3).
e. Leak around the
vaporizer filler
port
1. Remove the filler plug.
2. Verify that filler plug o-ring is not damaged.
Replace if damaged.
3. Verify that filler plug o-ring is not obstructed by
animal hair, etc.
4. Screw in the filler plug hand tight into the filler
port to form a tight seal.
7
Patient sleep level
seems too deep
a. Anesthetic
concentration is
set too high
Reduce concentration of anesthetic on the system
touch screen.
b. Vaporizer
malfunction
Contact Vetland or your local dealer for servicing.
8
Breathing bag is
overly distended
a. Pop-off valve is
closed
Open the pop-off valve
b. Flow rate is set
too high
Decrease flow rate by pressing the green circle at
the base of the flowmeter on the touch screen and
press the decrease button
T
until the desired flow
rate is displayed.
c. Scavenger
system suction is
set too low or is
occluded
Increase vacuum flow for scavenger suction or
install a scavenger interface valve vented to
atmosphere.
9
Gas is flowing , but
breathing bag does
not inflate
a. Breathing bag is
punctured
Replace breathing bag and mount the new bag
securely on the bag mount (Figure 1-3, 13)
b. Pop-off valve is
completely open
Turn the knob on the pop-off valve clockwise until
the bag starts to inflate.
c. Scavenger
system suction is
set too high
Decrease scavenger system suction.
or
Install a scavenger interface valve vented to
atmosphere.
Содержание EX3000
Страница 1: ...P N 410 125 EX3000 Veterinary Anesthesia System Operator s Manual ...
Страница 8: ......
Страница 20: ......
Страница 42: ......
Страница 56: ......
Страница 60: ......
Страница 67: ...Replacement Parts and Accessories 10 3 Figure 10 1 EX3000 Anesthesia System ...
Страница 69: ...Replacement Parts and Accessories 10 5 Figure 10 3 Optional Keyfill Vaporizer Fillers ...
Страница 70: ...10 6 Replacement Parts and Accessories Figure 10 4 410 105A Isoflurane Chamber Assembly ...
Страница 71: ...Replacement Parts and Accessories 10 7 Figure 10 5 410 106A Sevoflurane Chamber Assembly ...
Страница 72: ...10 8 Replacement Parts and Accessories Figure 10 6 410 102A Sub Chassis Assembly ...
Страница 73: ...Replacement Parts and Accessories 10 9 Figure 10 7 410 108A Circuit Board Assembly ...
Страница 74: ...10 10 Replacement Parts and Accessories Figure 10 8 410 100A Back Cover Assembly ...
Страница 75: ...Replacement Parts and Accessories 10 11 Figure 10 9 410 101A DISS Inlet Assembly ...
Страница 76: ...10 12 Replacement Parts and Accessories Figure 10 10 500 100A Pole Assembly ...
Страница 77: ...Replacement Parts and Accessories 10 13 Figure 10 11 510 570A APL Pop off Valve Assembly ...
Страница 78: ...10 14 Replacement Parts and Accessories Figure 10 12 510 103A Manifold Block Assembly ...
Страница 79: ...Replacement Parts and Accessories 10 15 Figure 10 13 510 101A Absorber Canister Assembly ...
Страница 80: ...10 16 Replacement Parts and Accessories Figure 10 14 510 129A Flush Valve Assembly ...
Страница 81: ...Replacement Parts and Accessories 10 17 Figure 10 15 510 104A Inspiration Expiration Valve Assembly ...
Страница 82: ......
Страница 83: ...Appendix A A 1 Appendix Figure A 1 EX3000 Machine Overview Diagram EX3000 Machine Overview Diagram ...
Страница 84: ...A 2 Appendix A Figure A 2 EX3000 Pneumatic Block Diagram EX3000 Pneumatic Block Diagram ...
Страница 85: ...Appendix A A 3 Figure A 3 EX3000 Electrical Block Diagram EX3000 Electrical Block Diagram ...
Страница 86: ...A 4 Appendix A Figure A 4 EX3000 Electrical Connections EX3000 Electrical Connections ...
Страница 87: ......