![Vetland EX3000 Скачать руководство пользователя страница 44](http://html.mh-extra.com/html/vetland/ex3000/ex3000_operators-manual_1010097044.webp)
6-2
Troubleshooting
Problem – Solution Matrix
Table 6-1 contains problems that may occur during operation of your EX3000 Anesthesia System and
their corrective actions. If you continue experiencing a problem after trying the suggested corrective
actions, please contact Vetland or your local dealer for advanced fault diagnostics.
Table 6-1 EX3000 Anesthesia System Problem – Solution Matrix
No.
Problem
Cause
Solution
1
No or low anesthetic
vapor output
a. Vaporizer is
functioning
properly, but
machine output is
not getting to
patient
Check breathing system components for leaks,
tears, holes, etc. Ensure the cone/mask makes a
good seal with the patient. Perform a System
Pressure Leak Test as specified below.
b. Anesthetic agent
reservoir is empty
Fill the reservoir with appropriate anesthetic drug.
c. Vaporizer is
turned off
1. Press
NONE
on the Agent Type Display area
(Figure 1-5, 13) of the touch screen.
2. Press the Increase (Up) Button
S
until the
name of the desired agent is displayed.
3. Press the Agent Concentration % area (Figure
1-5, 2) of the touch screen.
4. Press
S
on the touch screen until the desired
concentration is displayed.
d. Leak around the
vaporizer filler
port
1. Remove the filler plug.
2. Verify that filler plug o-ring is not damaged.
Replace if damaged.
3. Verify that filler plug o-ring is not obstructed by
animal hair, dirt, etc.
4. Screw in the filler plug hand tight into the filler
port to form a tight seal.
e. Vaporizer
malfunction –
internal fault
Contact Vetland or your local dealer for servicing.
2
Pop-off (APL) valve
knob is hard to turn
Pop-off valve
threads require
cleaning
Contact Vetland or your local dealer for servicing.
3
Needle on system
pressure gauge
(Figure 1-5, 3) does
not read zero (0) or is
fluctuating when all
gases are turned off
Internal fault
Contact Vetland or your local dealer for servicing.
4
Needle on system
pressure gauge
(Figure 1-5, 3)
indicates several
centimeters of
positive pressure
after exhalation has
ended
Constriction or
obstruction
downstream from
the pop-off (APL)
valve (Figure 6-1)
Disconnect the pop-off valve exhaust hose and
monitor the touch screen pressure gauge.
x
If the pressure
does not
return to zero,
contact Vetland or your local dealer.
x
If pressure
returns to zero
:
a. Check the exhaust hose for possible
obstruction due to moisture buildup or
debris.
b. Remove the obstruction.
c. Reconnect the hose to the pop-off valve.
Содержание EX3000
Страница 1: ...P N 410 125 EX3000 Veterinary Anesthesia System Operator s Manual ...
Страница 8: ......
Страница 20: ......
Страница 42: ......
Страница 56: ......
Страница 60: ......
Страница 67: ...Replacement Parts and Accessories 10 3 Figure 10 1 EX3000 Anesthesia System ...
Страница 69: ...Replacement Parts and Accessories 10 5 Figure 10 3 Optional Keyfill Vaporizer Fillers ...
Страница 70: ...10 6 Replacement Parts and Accessories Figure 10 4 410 105A Isoflurane Chamber Assembly ...
Страница 71: ...Replacement Parts and Accessories 10 7 Figure 10 5 410 106A Sevoflurane Chamber Assembly ...
Страница 72: ...10 8 Replacement Parts and Accessories Figure 10 6 410 102A Sub Chassis Assembly ...
Страница 73: ...Replacement Parts and Accessories 10 9 Figure 10 7 410 108A Circuit Board Assembly ...
Страница 74: ...10 10 Replacement Parts and Accessories Figure 10 8 410 100A Back Cover Assembly ...
Страница 75: ...Replacement Parts and Accessories 10 11 Figure 10 9 410 101A DISS Inlet Assembly ...
Страница 76: ...10 12 Replacement Parts and Accessories Figure 10 10 500 100A Pole Assembly ...
Страница 77: ...Replacement Parts and Accessories 10 13 Figure 10 11 510 570A APL Pop off Valve Assembly ...
Страница 78: ...10 14 Replacement Parts and Accessories Figure 10 12 510 103A Manifold Block Assembly ...
Страница 79: ...Replacement Parts and Accessories 10 15 Figure 10 13 510 101A Absorber Canister Assembly ...
Страница 80: ...10 16 Replacement Parts and Accessories Figure 10 14 510 129A Flush Valve Assembly ...
Страница 81: ...Replacement Parts and Accessories 10 17 Figure 10 15 510 104A Inspiration Expiration Valve Assembly ...
Страница 82: ......
Страница 83: ...Appendix A A 1 Appendix Figure A 1 EX3000 Machine Overview Diagram EX3000 Machine Overview Diagram ...
Страница 84: ...A 2 Appendix A Figure A 2 EX3000 Pneumatic Block Diagram EX3000 Pneumatic Block Diagram ...
Страница 85: ...Appendix A A 3 Figure A 3 EX3000 Electrical Block Diagram EX3000 Electrical Block Diagram ...
Страница 86: ...A 4 Appendix A Figure A 4 EX3000 Electrical Connections EX3000 Electrical Connections ...
Страница 87: ......