![Vetland EX3000 Скачать руководство пользователя страница 23](http://html.mh-extra.com/html/vetland/ex3000/ex3000_operators-manual_1010097023.webp)
Safety
2-3
Preventing System and Facility Damage
x
Immediately contain and clean up spilled anesthetic as recommended in the manufacturer’s
MSDS.
x
Isoflurane is very caustic and may dull the paint finish of the EX3000 if spillage occurs. If spillage
should occur, allow it to evaporate. Do not attempt to wipe it with a cloth.
Vetland
recommends
using a filler attachment to add anesthetic agents.
x
Do not place more than 7 pounds (3.2 kg) on top of the Mayo tray (if installed).
x
Follow all recommended system maintenance procedures specified in
Section 7 – Maintenance
.
x
Make sure the EX3000 is properly secured to the wall of the facility if used as a wall-mounted
system. Inappropriate fasteners may pull out and cause the system to fall to the floor.
x
Make sure electrical and gas supplies are within the required ratings.
Facility Environmental Health
x
Immediately contain and clean up any spilled anesthetic agent as recommended in the
manufacturer’s MSDS.
x
If the system is used in a confined space, ensure that there is adequate ventilation. If volatile
organic compound (VOC) emissions exceed safe limits, facility ventilation systems must be in
operation.
x
Dispose of any hazardous materials and items contaminated with hazardous materials in
accordance with applicable Federal, state, and local statutes and regulations, facility practice, and
practices recommended by the MSDS.
Alarms
The EX 3000 is equipped with audible and visual display alarms that alert the operator to potentially
unsafe conditions. Refer to
Section 6 – Troubleshooting
for alarm descriptions, possible causes, and
suggested solutions.
Emergency Shutdown
In the event of a major malfunction, shut down the EX3000 Anesthesia System as follows:
1. Press
STOP
on the system display panel for at least 2 seconds.
2. Turn OFF (
0
) the power switch on the back of the anesthesia machine.
3. Turn off the oxygen supply.
4. Unplug machine from the facility power source.
5. Disconnect the facility O
2
supply line (if used).
6. Switch the patient to a backup anesthesia delivery system and continue the medical procedure.
7. Contact
Vetland
or your local dealer for service.
Содержание EX3000
Страница 1: ...P N 410 125 EX3000 Veterinary Anesthesia System Operator s Manual ...
Страница 8: ......
Страница 20: ......
Страница 42: ......
Страница 56: ......
Страница 60: ......
Страница 67: ...Replacement Parts and Accessories 10 3 Figure 10 1 EX3000 Anesthesia System ...
Страница 69: ...Replacement Parts and Accessories 10 5 Figure 10 3 Optional Keyfill Vaporizer Fillers ...
Страница 70: ...10 6 Replacement Parts and Accessories Figure 10 4 410 105A Isoflurane Chamber Assembly ...
Страница 71: ...Replacement Parts and Accessories 10 7 Figure 10 5 410 106A Sevoflurane Chamber Assembly ...
Страница 72: ...10 8 Replacement Parts and Accessories Figure 10 6 410 102A Sub Chassis Assembly ...
Страница 73: ...Replacement Parts and Accessories 10 9 Figure 10 7 410 108A Circuit Board Assembly ...
Страница 74: ...10 10 Replacement Parts and Accessories Figure 10 8 410 100A Back Cover Assembly ...
Страница 75: ...Replacement Parts and Accessories 10 11 Figure 10 9 410 101A DISS Inlet Assembly ...
Страница 76: ...10 12 Replacement Parts and Accessories Figure 10 10 500 100A Pole Assembly ...
Страница 77: ...Replacement Parts and Accessories 10 13 Figure 10 11 510 570A APL Pop off Valve Assembly ...
Страница 78: ...10 14 Replacement Parts and Accessories Figure 10 12 510 103A Manifold Block Assembly ...
Страница 79: ...Replacement Parts and Accessories 10 15 Figure 10 13 510 101A Absorber Canister Assembly ...
Страница 80: ...10 16 Replacement Parts and Accessories Figure 10 14 510 129A Flush Valve Assembly ...
Страница 81: ...Replacement Parts and Accessories 10 17 Figure 10 15 510 104A Inspiration Expiration Valve Assembly ...
Страница 82: ......
Страница 83: ...Appendix A A 1 Appendix Figure A 1 EX3000 Machine Overview Diagram EX3000 Machine Overview Diagram ...
Страница 84: ...A 2 Appendix A Figure A 2 EX3000 Pneumatic Block Diagram EX3000 Pneumatic Block Diagram ...
Страница 85: ...Appendix A A 3 Figure A 3 EX3000 Electrical Block Diagram EX3000 Electrical Block Diagram ...
Страница 86: ...A 4 Appendix A Figure A 4 EX3000 Electrical Connections EX3000 Electrical Connections ...
Страница 87: ......