At the Actuator Cabinet
1.
Installer #2 retracts each of the cables back into the cabinet in short
increments, while at the same time Installer #1 moves the source carrier
cover closer to the source holder.
2.
Continue the progression of retracting the cables until Installer #1 has the
source carrier cover pressed fully against the source holder.
3.
Fasten the source carrier cover to the holder using the 3/4
" socket and
ratchet
and the same hex bolts that secured the shipping cover to the
source holder when originally shipped.
Insert Sources and Establish Elevations
Tools needed for this part of installing the SHLM MI include:
•
1/8
"
hex key
1.
Remove the combination lock from the shutter.
2.
Pull the shutter outward until it locks into the detent.
3.
Return to the actuator cabinet with the 1/8
" hex key.
4.
To establish proper source carrier elevation within process vessel:
5.
Release the source carrier you wish to insert by releasing it from its cable
latch inside the actuator cabinet.
6.
Confirm the insert cable moves freely through the detent plug by making
sure set screws in the plug are not tightened against the cable.
7.
Begin inserting the source carrier into the vessel by pushing the cable
downward and into flex conduit.
The task of retracting the cables requires the installer to pull a cable
back into the cabinet with their right hand, while using their left hand
to spin the full cable coil in a direction that will capture the cable
within the guide lanes.
You may find it necessary to manually rotate the cable bundle within
the cable guides to keep the cable moving easily during the
insertion process.
!
NOTE
!
NOTE
Содержание SHLM MI4
Страница 1: ...Operating Instructions SHLM MI4 Source Holder Document ID 48642 Nuclear...
Страница 2: ......
Страница 4: ...4 48642 US 150527 NOTES...
Страница 8: ...8 48642 US 150527 NOTES...
Страница 12: ...12 48642 US 150527 NOTES...
Страница 16: ...16 48642 US 150527 NOTES...
Страница 26: ...26 48642 US 150527 NOTES...
Страница 31: ...48642 US 150527 31 Figure 4 2 SHLM MI4 Preferred Density and Point Level Orientation...
Страница 43: ...48642 US 150527 43 Figure 5 4 Mounting Frame without Cabinet...
Страница 44: ...44 48642 US 150527 NOTES...
Страница 64: ...64 48642 US 150527 NOTES...
Страница 70: ...70 48642 US 150527 NOTES...
Страница 71: ......