32
48642-US-150527
Source Interference
When multiple adjacent pipes, vessels, or conveyors have nuclear gauges, you
must consider the orientation of the source beams so each gauge senses
radiation only from its appropriate source. Maximize the distance between the
source holders when mounting them on adjacent vessels. Make sure to maintain
the appropriate source to detector distance as illustrated in
Figure 4.3
.
Figure 4.3
Source/Detector Arrangement for Adjacent Vessels
1. Radiation Beam
2. Detector
3. Source Holders
4.5 Temperature
Your source holder may have components that are sensitive to temperature
extremes. Avoid exposure to temperatures beyond the specified limits.
The
maximum operating temperature for the vessel is + 260 °C
(+ 500 °F). Consult the
factory if this temperature is exceeded.
Ambient Temperature
You may use the source holder in continuous ambient temperature conditions from
- 40 °C … + 60 °C (-40 °F … + 140 °F). Minimize direct exposure to sunlight when
ambient temperatures 40 °C
(+ 105 °F).
2
1
1
3
2
Содержание SHLM MI4
Страница 1: ...Operating Instructions SHLM MI4 Source Holder Document ID 48642 Nuclear...
Страница 2: ......
Страница 4: ...4 48642 US 150527 NOTES...
Страница 8: ...8 48642 US 150527 NOTES...
Страница 12: ...12 48642 US 150527 NOTES...
Страница 16: ...16 48642 US 150527 NOTES...
Страница 26: ...26 48642 US 150527 NOTES...
Страница 31: ...48642 US 150527 31 Figure 4 2 SHLM MI4 Preferred Density and Point Level Orientation...
Страница 43: ...48642 US 150527 43 Figure 5 4 Mounting Frame without Cabinet...
Страница 44: ...44 48642 US 150527 NOTES...
Страница 64: ...64 48642 US 150527 NOTES...
Страница 70: ...70 48642 US 150527 NOTES...
Страница 71: ......