Insert the Carrier Module with the MTT module inside the slot.
Insert the FAT32 USB memory stick containing the software package into one of the USB ports.
Turn the RXT-1200
ON.
Go to
>Utilities >M.Upgrade.
Once the upgrade image is recognized, confirm the versions and tap on the
Upgrade
button.
The firmware update starts.
It will turn itself off when finished.
MTT Auto-Update:
If an older MTT module is inserted into an updated RXT-Carrier, the RXT may start the MTT firmware upgrade process.
A message will be presented on the screen indicating the version discrepancy.
Press
Yes
to update the MTT module’s firmware.
This process doesn’t require the USB memory stick.
MTT modules updated to work with the new RXT-1200 may no longer be fully compatible with older SunSet MTT chassis or RxT-
1000 platforms
.
RXT1200_Platform_Manual_RevB01
Page 52 of 117
Содержание RXT-1200
Страница 1: ......
Страница 17: ...Go back to TOC RXT1200_Platform_Manual_RevB01 Page 17 of 117...
Страница 76: ...Result tab Tree structure file storage RXT1200_Platform_Manual_RevB01 Page 76 of 117...
Страница 113: ...GNSS Start Menu GNSS Start GNSS Results RXT1200_Platform_Manual_RevB01 Page 113 of 117...