TECHNICAL INFORMATION
65
87-900-855-01 (B)
•
Choose the desired selection by pressing the
CYCLE push-button.
F R O N T / R E M O T E / S E R
S E L E C T I O N : F R O N T
Instead of SER, one of configured RS 232, RS 485, RS
422 ports will be displayed.
•
Press the PUMP CURRENT push-button to enter
the value and, if the serial option is installed, the
display shows:
R S 2 3 2
B A U D
R A T E
X X X X
where:
XXXX
= means 600, 1200, 2400, 4800, 9600
baud rate for the host computer or printer
communication. The controller is factory-set for 9600
baud rate operation.
Instead of RS 232, one of configured RS 232,
RS 485, RS 422 ports will be displayed.
•
Select the desired value by pressing the CYCLE
NUMBER.
R S 2 3 2
B A U D
R A T E
9 6 0 0
Instead of RS 232, one of configured RS 232,
RS 485, RS 422 ports will be displayed.
•
Enter the value by pressing the PUMP CURRENT
push-button and, if the serial option is installed, the
display shows:
R S 2 3 2
H O S T / P R I N T
S E L E C T I O N :
X X X X X
where:
XXXXX
= means HOST or PRINT.
This screen will only be displayed if the RS 232
interface is selected.