TECHNICAL INFORMATION
60
87-900-855-01 (B)
Window-Based Protocol
The following table, valid for the RS 232, RS 422 and RS 485 interfaces, describes each single window used in the
protocol.
WIN
R
W
T
DESCRIPTION
000
X
X
L
START/STOP
001
X
X
L
LOW SPEED ON/OFF
100
X
X
L
SOFT START YES/NO
101
X
X
L
DEAT TIME YES/NO
102
X
X
L
WATER COOLING YES/NO
103
X
X
N
SPEED THRESHOLD
104
X
X
N
RUN UP TIME (0÷359.999 sec)
105
X
X
N
LOW SPEED ADJUST
106
X
X
N
HIGH SPEED ADJUST
107
X
X
N
MODE (0, 1, 2) [FRONT, REMOTE, SERIAL]
108
X
X
N
BAUD_RATE (0-4) [600, 1200, 2400, 4800, 9600]
109
X
L
PUMP LIFE RESET (TYPE “ON” TO RESET)
120
X
X
L
EXTERNAL INPUT READ ENABLE [YES/NO]
121
X
X
L
HEATER BAND ENABLE [YES/NO]
200
X
N
CURRENT [ A]
201
X
N
VOLTAGE [ V]
202
X
N
POWER [ W]
203
X
N
EXCITEMENT FREQUENCY [ Hz]
205
X
N
PUMP STATE (0÷6) [STOP, WAITING INTERLOCK, STARTING, NORMAL,
HIGH LOAD, FAILURE,APPROACHING]
206
X
N
ERROR CODE (0÷8) [NO ERROR, OVERVOLTAGE, SHORT CIRCUIT, CHECK
CONNECTION, TOO HIGH LOAD, OVERRIDE, PUMP
OVERTEMP, CONTROLLER OVERTEMP, OIL LEVEL AT MIN]
207
X
L
STATE R1 ON/OFF
208
X
L
STATE R2 ON/OFF
209
X
N
UPPER BEARING TEMPERATURE (0÷90) [ °C]
210
X
N
LOWER BEARING TEMPERATURE (0÷90) [ °C]
217
X
N
MOTOR TEMPERATURE (0÷90) [ °C]
218
X
N
ROTATION SPEED [Krpm]
219
X
N
EXTERNAL INPUT [V]
300
X
N
CYCLE TIME (0-999.999) IN MINUTES
301
X
N
CYCLE NUMBER (0- 65.535) IN COUNTS
302
X
N
PUMP LIFE (0-999.999) IN HOURS
400
X
A
CRC PROGRAM LISTING
402
X
A
CRC PARAMETER LISTING
WIN
= Window
T
= Type:
R
=
Read
L
= Logical
W
=
Write
N
= Numeric
A
= Alphanumeric
Содержание Turbo-V 6000
Страница 2: ...Turbo V 6000 Controller...
Страница 3: ......
Страница 4: ......
Страница 80: ...TECHNICAL INFORMATION 74 87 900 855 01 B...