Hydraulic connections
Hydraulic connections are located at the front of the ProStar 335
detector.
The only line installed by the user where dead volume and low
holdup are critical is the line from the column exit to the flowcell
inlet port. This line should be as short as possible.
The ProStar 335 detector can be fitted with any one of four
flowcells (ordered separately). Each one has an inlet and outlet
connection and quartz optics cell window. Your chosen flowcell
type is packed internally in the detector. The four flowcell types
are outlined in Table 2.
Table 2 Compatible flowcells
Flowcell type
Flowcell P/N
Pathlength*
Column ID
Analytical
02-101818-00
9 mm x 0 mm
2 to 8 mm
Preparative
02-101819-00
9 mm x 1 mm
4 to 76 mm
Super Prep
02-101820-00
4 mm x 0.15 mm
8 to 152 mm
Microbore
02-101821-00
4 mm x 0 mm
1 to 4 mm
* A pathlength of a mm x b mm means that the sample light path
has a pathlength of a mm, and the reference light path has a
pathlength of b mm. A reference light path of zero means there is
no fluid in the cell — air acts as the reference.