Relay output
For time programming external events, a contact closure Relay
output is available. To configure the Relay output, install the
optional Relay output board (part number 02-101875-90) into the
J4 receptacle. Pin designations are shown below.
Figure 5 Pin designation for J4
There are four general-purpose output relays and one dedicated
Peak relay. Each output uses a DIP relay that is capable of
handling 500 mA of contact current. At reset or power up, the
output relay contacts are set to the default parameters (open).
After loading a method they will be set as defined in the method’s
time=0 parameters.