Replacing the Deuterium (UV) lamp
Your new replacement lamp is delivered to you pre-tested and
aligned in a mounting bracket. It is designed for direct
replacement and requires no alignment procedures after it has
been installed.
Work through the following steps for removal of the old lamp, and
installation of the new one.
WARNING:
To avoid electric shock and eye damage, turn the main power switch to OFF and
disconnect the power cord before starting the lamp replacement procedure.
WARNING:
Let the lamp housing cool before removing the lamp.
NOTE:
You should wear gloves when replacing the lamp to prevent any oil or
other material from getting on the lamp glass envelope. It is important
that you never touch the glass on the lamp with your bare hands, as this
can cause the lamp to fail prematurely or even crack.