46
03-914648-00:7
Page two, on GCs purchased with detector EFC parameters, is accessed by
pressing the PAGE DOWN key.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Make-up Flow (ml/min): 25 H2 Flow (ml/min): 30
Air Flow (ml/min): 300
FRONT FID ADJUSTMENTS, Page 2 of 2
Make-up Flow (ml/min)
25
H
2
Flow (ml/min)
30
Air Flow (ml/min)
300
The upper portion of the screen indicates the actual flow rates for the various
detector gases. The method parameters, default settings and range are:
Default
Range
Make-up Flow
25 ml/min
0 - 50 ml/min
H
2
Flow
30 ml/min
0 - 50 ml/min
Air Flow
300 ml/min
0 - 500 ml/min
METHANIZER
A methanizer is an option that converts carbon monoxide and carbon
dioxide to methane. It is associated with a specific FID detector;
consequently, it is accessed in the DETECTOR section. A methanizer is
presented as one of the detector options in the detector menu screen and is
identified by location as “front”, “middle, and “rear”. Below is a sample
screen for a Methanizer.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Set (°C): 380 Actual (°C): 380
REAR METHANIZER
Methanizer Oven
ON
Temperature
380