20
03-914648-00:7
or the CURSOR UP/DOWN keys can be used to scroll one line at a time.
Status parameters appear in the same order as the sample flows through the
system, i.e., Injector, Flow / Pressure, Column Oven, Detector, and Output.
Local Automation status indicates the current automation status if a series of
runs has been started from the local keyboard of the 3800.
The status page also indicates whether a component has failed to reach its
setpoint or has a fault. If a setpoint is not reached, the message NOT
READY is displayed next to the component, while if a fault is present the
message FAULTED appears. The Column Oven field also indicates when it
is stabilizing.
Miscellaneous
Setup
Parameters
•
Is Ready-In a closed contact (Yes/No). Allows you to specify what switch state
corresponds to External Device Ready.
•
Enable check for (Front/Middle/Rear) FID flameout (Yes/No).
If you have an FID electrometer installed, this field allows the GC to ignore
flameout test results. This could be used if you have a very low background
system, or when the FID electrometer is used to collect signal from another low-
background detector, such as a PID.