CP-3800 GC Operator’s Manual
167
Single Method, Non-Addressable Automation
If an addressable device such as an 8400 AutoSampler is not present, the
3800 defaults to the non-addressable mode on power up. In Single Method
Automation the same 3800 method can be run up to 999 times.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
SINGLE METHOD AUTOMATION
Method: Method 1
Injections: 1
Select from one of the available eight 3800 methods using the increment or
decrement keys.
Injections
refers to how many times this method should be
run. The allowable entries are 1 to 999.
To start a Single Method Automation, press the Select / Edit key in the
Automation section of the 3800 keyboard, then select Single Method as the
automation choice. Press the 3800 Start key to start automation. Note that if
the method specified in Single Method Automation is different from the
currently active 3800 method, this may result in a delay of several minutes
before the first analytical run starts.
Method Sequence, Non-Addressable Automation
Method Sequence Automation allows running an automation sequence
where the method can be changed from run to run. The sequence of
methods is built by adding lines to the automation table. In addition, a
method sequence automation table may be run multiple times. The method
sequence automation table can consist of up to 25 lines. Lines are added by
pressing the New Line key on the 3800 keyboard. Page 2 of the Method
Sequence allows repetition of the sequence up to 999 times.