Model 990 dCLD II Component Leak Detector
3-13
REJECT SET POINTS
The REJECT SET POINTS are used to alert the operator that a mea-
sured leak rate has exceeded the set value. There are four set points
available. Each set point must be enabled before they are opera-
tional.
To set the value of the REJECT SET POINTS, issue the
INIT-
n
REJECT<CR> command, where
n
is equal to 1 through 4. If
the optional front control panel is installed, enter the value through
the Reject and Audio Set Points screen (Section 4.6.2 “Reject and
Audio Set Points” on page 4-14).
ENABLE AUDIO
REJECT
AUDIO REJECT must be enabled before the function is operational
and reacts based upon its set point value.
To ENABLE the AUDIO set point, issue the ENABLE-AREJECT<CR>
command or, if the optional front control panel is installed, enable
AUDIO by setting the ON/OFF button located to the right of the
audio set point field to ON in the Reject and Audio Set Points screen
(Section 4.6.2 “Reject and Audio Set Points” on page 4-14).
To DISABLE the AUDIO set point, issue the DISABLE-AREJECT<CR>
command or, if the optional front control panel is installed, disable
AUDIO by setting the ON/OFF button to OFF in the Reject and
Audio Set Points screen (Section 4.6.2 “Reject and Audio Set
Points” on page 4-14).
AUDIO SET POINT
The AUDIO SET POINT control is used to activate the audio tone
from the optional front control panel when the measured leak rate
value exceeds the threshold value (leak rate pass/fail specification)
set by the operator. The audio tone frequency increases as the size
of the measured leak increases beyond the set point. As stated
above, you must enable AUDIO REJECT before the function is
operational. If AUDIO REJECT is disabled, the audio tone fre-
quency will increase from zero leak rate instead of from the audio
reject set point.
The audio volume control is located on the optional front control
panel (see Figure 4-1 on page 4-1). The sound emanates from the
rear of the optional front control panel.
To set the Audio Reject set point, issue the INIT-AREJECT<CR>
command or, if the optional front control panel is installed, set the
audio reject set point value in the Reject and Audio Set Points screen
(Section 4.6.2 “Reject and Audio Set Points” on page 4-14).
Содержание 990 dCLD II
Страница 4: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank D R A F T 4 2 0 0 7 ...
Страница 12: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 50: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 128: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 146: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 160: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 161: ......