Model 990 dCLD II Component Leak Detector
2-8
2.4 Electronic Considerations
This section includes information necessary to interface the 990 dCLD II with a PLC or a
host computer using either serial or parallel communications. This includes connection via
J3 Discrete I/O, J4 BCD Outputs and J6 Host Serial Port.
2.4.1 Discrete I/O J3 Inputs
The discrete I/O is used to control operations via a PLC. Table 2-7 provides the input con-
nections. For detailed information see Table B-1 on page B-1. For a cross reference of the
function in other modes of operation, see Table 2-11 on page 2-15.
J3 Level inputs are opto-isolated, 5 to 24 VDC 3600 Ohm resistive load. J3 Momentary
inputs are opto-isolated, 5 to 24 VDC 3600 Ohm resistive load, and require 200 ms mini-
mum pulse width. Figure 2-4 shows the J3 input circuit sketch.