3-1
Model 990 dCLD II Component Leak Detector
Section 3. Setup and Operations via the
Rear Panel
3.1 Physical Setup
Figure 3-1 on page 3-3 shows the system block diagram for the 990 dCLD II. Information
regarding the physical setup is discussed below.
3.1.1 Control Unit
The Control Unit is designed to be mounted in a 19-inch rack cabinet. Air is required to
flow through the control unit entering through the two filtered air intakes at the front of the
unit and exiting at the rear of the unit through a fan. Locate the control unit in a way that
allows easy viewing of the front LEDs and easy access to the rear panel connections.
3.1.2 Turbo Spec Tube Assembly
The Turbo Spec Tube assembly can be mounted in any orientation. The spectrometer tube
must be isolated from strong magnetic fields such as motors, transformers, and magnets.
Steel structures and steel objects should also be kept at least two inches away from the clos-
est edge of the spectrometer tube. During normal operation, the spectrometer tube feels
quite warm. A continuously running fan at the end of the high vacuum pump directs cool-
ing air toward the end of the pump. The pump operates between 5 °C and 35 °C ambient
temperature.
Содержание 990 dCLD II
Страница 4: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank D R A F T 4 2 0 0 7 ...
Страница 12: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 50: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 128: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 146: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 160: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 161: ......