Model 990 dCLD II Component Leak Detector
1-8
System I/O Capability
J3 Inputs
Opto-isolated, 5 to 24 VDC 3600 Ohm resistive
load.
J3 Outputs
Emitter follower with 10 Ohm series resistor,
14 mA max drive current, 24 VDC max.
J3 Momentary Inputs
Opto-isolated, 5 to 24 VDC 3600 Ohm resistive
load, requires 200 ms per min pulse width.
J3 Analog Leak Rate Output
0 to 10 V per decade linear; 1, 2 or 3 V per
decade logarithmic. Note: 3 V per decade
logarithmic has limited use on the highest
decade.
J4 BCD Leak Rate Output
Emitter follower with 10 Ohm series resistor,
14 mA max drive current, 24 VDC max. Note:
This output is valid only when the leak rate output
pulse is low. The leak rate is updated every
50 ms.
J6 Host Serial Port
9600 Baud, No parity, 8 Bits, 1 Stop Bit,
interrupt-driven RS-232 port for connection to
a host computer.
J12 Diagnostic Port
9600 Baud, No parity, 8 Bits, 1 Stop Bit,
configured as DTE (Data Terminal Equipment).
Use null modem cable when connecting to a host
computer.
Содержание 990 dCLD II
Страница 4: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank D R A F T 4 2 0 0 7 ...
Страница 12: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 50: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 128: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 146: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 160: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 161: ......