Model 990 dCLD II Component Leak Detector
3-11
NORMAL
CALIBRATION
The NORMAL calibration routine performs a thorough tuning pro-
cess and a system gain adjustment (calibration). The full tuning pro-
cess involves independently scanning the ion source chamber
voltage and the variable focus voltage, and optimizing each value
for a maximum helium signal. Once the tuning process is com-
plete, the preamplifier gain is adjusted to bring the leak detector
into calibration.
The SPEC TUBE LED on the front of the control unit flashes while
calibration is taking place, the Discrete I/O J3/22 is active, and the
response to the inquiry ?STATUS<CR> includes CAL. If the optional
front control panel is installed, the calibrate indicator LED lights
while calibration is taking place.
To set the mode to NORMAL, issue the 0 INIT-QUICK-CAL<CR>
command or, if the optional front control panel is installed, select
NORM from the CAL LEAK SET-UP screen (Section 4.6.1
“Calibrated Leak Set-Up” on page 4-13).
FAST
CALIBRATION
During FAST calibration, the leak detector compares the leak rate
signal with the calibrated leak value and, if the gain adjustment
required to bring the system into calibration is within the allowable
limits. The full tuning operation is omitted.
The SPEC TUBE LED on the front of the control unit flashes while
calibration is taking place, the Discrete I/O J3/22 is active, and the
response to the inquiry ?STATUS<CR> includes CAL. If the optional
front control panel is installed, the calibrate indicator LED lights
while calibration is taking place.
If the calibration parameters are set to INTERNAL, the system notes
the leak rate, turns off the leak and ZEROs the system. Then the sys-
tem performs a NORMAL or FAST calibration as described above.
The SPEC TUBE LED on the front of the control unit flashes while
calibration is taking place, the Discrete I/O J3/22 is active, and the
response to the inquiry ?STATUS<CR> includes CAL. If the optional
front control panel is installed, the calibrate indicator LED lights
while calibration is taking place.
If the calibration parameters are set to EXTERNAL, the system per-
forms a NORMAL or FAST calibration as described above.
Содержание 990 dCLD II
Страница 4: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank D R A F T 4 2 0 0 7 ...
Страница 12: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 50: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 128: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 146: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 160: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 161: ......