![background image](http://html1.mh-extra.com/html/varian/969-8988/969-8988_instruction-manual_854644033.webp)
TECHNICAL INFORMATION
91 87-900-986-01 (
C
)
Pressing the COUNTERS key, the following is
obtained:
C Y C L E
X X X X
T
I
M E
X X X X X m
Turbo-V 81-AG
LS
AG
COUNTERS
increase
decrease
LOW SPEED
menu
VACUUM TECHNOLOGIES
MEASURES
next
previous
START/STOP
R2
R1
Where
CYCLE
XXXX
Number of start/stop cycles
TIME
XXXXXm
Time of last cycle (minutes)
Pressing again:
P U M P
L
I
F E
X X X X X X h
Where:
PUMP LIFE
XXXXXX
total operating time (hours)of the
pump
Programming
Configuration menu
To access the configuration menu, press the
CO MEASURES buttons at the same time
for at least 2 sec.
Turbo-V 81-AG
LS
AG
COUNTERS
increase
decrease
LOW SPEED
menu
VACUUM TECHNOLOGIES
MEASURES
next
previous
START/STOP
R2
R1
In programming mode, the meaning of the 4 buttons
changes as follows:
–
COUNTERS/LOW SPEED
become “
INCREASE
and
DECREASE
” and make it possible to select the
next or previous value of a parameter (in the case
of compulsory parameters) or to increase/decrease
the individual digit of a parameter (in the case of
numeric parameters)
–
The “
MEASURES
and
START/STOP
” buttons
become “
NEXT
and
PREVIOUS
” and make it
possible to select the next parameter (or next digit
of a numeric parameter) or previous parameter.
Switching from one parameter to the next (using the
“
NEXT”
key) indicates confirmation of any changes
made to the specific parameter. If the changes cannot
be accepted by the controller, the “
Data out of limits
”
error message is displayed, the changes are cancelled
and the value of the parameter prior to the modification
is reproposed.