TECHNICAL INFORMATION

93 87-900-986-01 (

C

)

 

VENT VALVE MODE:

 makes it possible to select

driving of the vent valve

-

AUTO TIME: the valve is controlled by the
controller

-

 

MANUAL: the valve is controlled by the serial
port.

-

 

AUTO PUMP SPEED: the valve is controlled
by the controller its king into account the
pump’s deceleration.

 

VENT VALVE DELAY

: Sets the delay, expressed

in seconds, between stopping of the pump and
opening of the vent valve.

 

VENT VALVE OPEN TIME:

 sets the time,

expressed in seconds, of opening of the vent
valve. Enter 0 if the valve is to be kept open
indefinitely (until the next start).

 

LOAD GAS TYPE:

 selects the type of gas used in

the process. Therefore, the controller limits the
power supplied to the pump according to the
setting of this parameter.

 

WATER COOLING:

 indicates whether the pump is

cooled with water or not. The  controller limits the
power supplied to the pump according to this
setting.

 

ACTIVE STOP:

 enables (YES) or disables (NO)

active braking.

 

STOP SPEED READING:

 enables (YES) or

disables (NO) the pump rotational frequency
reading function also during the slowing phase that
follows a stop command

INPUT/OUTPUT menu

This menu contains parameters that determine the
operating mode of the inputs and outputs of the
controller.

START MODE 

START - INTERLOCK 

START/STOP LEVEL

START/STOP EDGE 

INTERLOCK TYPE 

CONTINUOUS  

IMPULSIVE    

ANALOG OUT  TYPE

LCD CONTRAST

FREQUENCY    

0  

POWER      

CURRENT     

7     

PRESSURE    

START OUT MODE 

RUNNING   

STARTING  

SPEED DISPLAY

KRPM     

Hz   

Flow  3/6

IMPUT / OPTPUT

F. 1

 

START MODE:

 Permits selection of one of the

three types of functioning of the START/STOP and
INTERLOCK inputs:

-

START/INTERLOCK: The pump is started if
the START/STOP input is activated and is
stopped if the START/STOP input is de-
activated. The interlock input must be
constantly active (INTERLOCK TYPE
parameter = continuous) or may be active only
at the time of starting (INTERLOCK TYPE
parameter = impulsive)

-

START/STOP LEVEL The START/STOP input
acts as only START command (active on the
level) The INTERLOCK input acts as STOP
command (active on the level)
In the case of concurrent activation of the two
inputs, the STOP command takes priority

-

START/STOP EDGE The START/STOP input
acts as START command (active on the edge)
The INTERLOCK input acts as STOP
command (active on the edge).

 

INTERLOCK TYPE:

  This parameter is displayed

only when the previous parameter (START MODE)
is equal to START/INTERLOCK and permits
selection of one of the following two operating
modes of the INTERLOCK input:
-

CONTINUOUS: the INTERLOCK input must
be active continuously so that the pump can
be driven by the controller. If, with the
controller in start, the input INTERLOCK is de-
activated, the controller switches to WAITING
INTERLOCK status and interrupts driving of
the pump

-

IMPULSIVE: it is sufficient for the input to be
active when the START command is sent to
the pump. Subsequently, the input can also be
de-activated. If  the interlock is not active when
the pump is started, the controller switches too
WAITING INTERLOCK status.

 

ANALOG OUT TYPE

: permits selection of the type

of signal to be sent to the programmable analogue
output. One of the following signals can be
selected:
-

FREQUENCY: the drive frequency of the
pump is shown

-

POWER: the power supplied to the pump is
shown

-

 

TEMPERATURE: the temperature of the pump
is shown

-

 

PRESSURE: the pressure read by the gauge
is shown.

 

START OUT MODE:

 Selects the operating mode

of the START output. Possible options:

-

 

RUNNING: the output is active when the pump
is driven

-

 

STARTING: the output is active only during
acceleration ramp-up.

 

SPEED OUT MODE (KRPM/Hz):

 permits selection

of the unit of measurement of the speed of rotation
shown on  the display of the controller.
This setting acts only on the measurement shown
on the display. Possible options:

-

 KRPM

-

 Hz

 

LCD CONTRAST

Содержание 969-8988

Страница 1: ...TRU ES NL BEDRIJFSHANDLEIDING DK ISTRUKSTIONSBOG S BRUKSANVISNING N INSTRUKSJON MANUAL FIN OHJEK SIKIRJA GR FELHASZN L I K ZIK NYV PL PODRECZNIK INSTRUKCJI CZ N VOD K POU IT SK N VOD NA OBSLUHU SLO PR...

Страница 2: ...Turbo V 81 AG Rack Controller...

Страница 3: ......

Страница 4: ......

Страница 5: ...S CRHSEWS 41 HASZN LATI UTAS T S 45 INSTRUKCJA UZYTKOWANIA 49 P RU KA K POU IT 53 N VOD K POU ITIU 57 NAVODILA ZA UPORABO 61 INSTRUCTIONS FOR USE 65 TECHNICAL INFORMATION 69 TURBO V 81 RACK CONTROLLER...

Страница 6: ...MODE menu 92 PUMP SETTING menu 92 INPUT OUTPUT menu 93 GAUGE menu 94 SERIAL menu 95 HOW TO USE BY REMOTE I O 96 HOW TO USE IN SERIAL MODE 96 PROFIBUS OPTION 97 General 97 Installation 97 Configuration...

Страница 7: ...ee the safety of the operator when using the equipment Detailed information is supplied in the appendix Technical Information This manual uses the following standard protocol The warning messages are...

Страница 8: ...is steady Never operate the Turbo V pump if the pump inlet is not connected to the system or blanked off NOTE The input signal J1 connector should be left in position including the shipping links if...

Страница 9: ...ific national standards DISPOSAL Meaning of the WEEE logo found in labels The following symbol is applied in accordance with the EC WEEE Waste Electrical and Electronic Equipment Directive This symbol...

Страница 10: ...MP The controller enviroment temperature exceeds 70 C Or The controller s radiator temperature is above 60 C Wait until the temperature decrease below threshold value Press the START push button twice...

Страница 11: ...vides and acquires the pressure of the wide range gauge Remote I Os compatible with the previous version Navigator default serial compatible with the previous RS 232 and 485 version Speed reading afte...

Страница 12: ...3 EN 61000 6 2 EN 61000 4 2 3 4 5 6 11 8 Weight 1 7 Kg 3 2 lbs Pollution Degree 2 Installation Category II CONTROLLER OUTLINE The outline dimensions for the Turbo V 81 AG Rack controllers are shown i...

Страница 13: ...ou already have fan installed use the available adapter cable to connect to the pump Fan Connector J7 Vent Vent Connector This is dedicated 24 Vdc connector to control the optional vent valve see orde...

Страница 14: ...Low Speed Contact open High Speed Optically isolated Vlow max 10v Imax 5mA Active only in Remote Mode operation 3 com Interlock Contact closed Active Contact open Inactive If active allows the pump r...

Страница 15: ...ins before measuring voltage A STOP command by front Panel or serial or Start stop input reset the FAIL status 13 6 Start Configurable output Default setting win 177 0 24V pump in RAMP state 0V other...

Страница 16: ...Can be related to frequency power pump temperature linear or pressure logarithmic The setting can be done by front panel or serial line win 111 0V 0Hz 0W 0 C 10V 963Hz 150W 100 C See following chart f...

Страница 17: ...01 C Diagram 3 R1 or R2 related to time Diagram 4 R1 or R2 related to Normal status Diagram 5 R1 or R2 related to pressure reading if gauge connected V Volt R1 R2 Stop Waiting Interlock Ramp Normal B...

Страница 18: ...grammable analog output related to frquency Diagram 2 Programmable analog output related to power Diagram 3 Programmable analog output related to pump temperature Frequency Hz 1350 V Volt Output Volta...

Страница 19: ...ONNECTOR This connector is available to connect EyeSys Mini IMG gauge with it s extention cable see orderable parts table The pressure can be read by dispaly or by serial line win 224 or by REMOTE I O...

Страница 20: ...can be activated by Front Panel or serial line Win 107 167 Active Stop brakes the pump using the motor The Stop Speed reading reads the pump speed after a stop command SSR AS Diagram 5 J2 SERIAL CONN...

Страница 21: ...e used as well with 232 or 485 media NOTE Please use Window protocol for new development LETTER PROTOCOL DESCRIPTION The default protocol is Window Communication format 8 data bit no parity 1 stop bit...

Страница 22: ...INTERLOCK 2 STARTING 3 NORMAL OPERATION 4 5 HIGH LOAD 6 FAILURE 7 APPROACHING LOW SPEED 2 5 Cycle Time 6 9 Pump Life 10 11 Pump Temperature 12 Current 13 Voltage 14 17 Frequency 18 19 Cycle 20 R1 Sta...

Страница 23: ...4800 9600 programmable Communication Protocol The communication protocol is a MASTER SLAVE type where Host MASTER Controller SLAVE The communication is performed in the following way 1 the host MASTER...

Страница 24: ...d instruction of a logic window Numeric 6 bytes after a read instruction of a numeric window Alphanumeric 10 bytes after a read instruction of an alphanumeric window ACK 1 byte 0x6 the command executi...

Страница 25: ...mmand READ SERIAL TYPE Source PC Destination Controller with address 3 in 485 mode 02 83 35 30 34 30 03 38 31 STX ADDR WINDOW RD ETX CRC Source Controller Destination PC 02 83 35 30 34 30 31 03 42 30...

Страница 26: ...bit 0 2 sec 155 R N Power limit applied Read the maximun allowable power watt 157 R W N Gas load type Select the gas load to the pump N2 1 Av 161 R W N Pressure Reading Connection Factor See GAUGE CON...

Страница 27: ...fe in hours zeroed by the reset command 0 to 999999 320 to 399 Reserved to Varian service 400 R A CRC EPROM QE QE8XXXX where XXXX are variable 402 R A CRC Param PA PA8XXXX where XXXX are variable 404...

Страница 28: ...ntroller is set to FRONT mode error message is returned If it is set to REMOTE mode the COMMAND LOCKED IN REMOTE error message is displayed while if it is in SERIAL mode the COMMAND LOCKED IN SERIAL M...

Страница 29: ...R M A L X X X X H z I X X X A V X X X V Where NORMAL indicates the operating regime of the pump XXXHz see description Pressing once Measures key I X XXA indicates the current absorbed by the pump V X...

Страница 30: ...ort with address YYYY Black rectangles as shown below may be displayed at any time in the first and last box of the bottom line of the display opposite the LS AG R1 R2 captions The meaning of these re...

Страница 31: ...OP R2 R1 This screen page will be visible throughout the acceleration phase the pump The increasing value of the rotation frequency will be displayed in the boxes marked X the user may decide to expre...

Страница 32: ...plied by the controller to the pump at that precise moment V is the voltage at which the controller is powering the pump at that moment Pressing the MEASURES key three times with the GAUGE connected a...

Страница 33: ...S MEASURES next previous START STOP R2 R1 In programming mode the meaning of the 4 buttons changes as follows COUNTERS LOW SPEED become INCREASE and DECREASE and make it possible to select the next or...

Страница 34: ...dicated below FRONT accepts commands from the front panel REMOTE accepts commands from the remote input connector SERIAL accepts commands from the serial interface or Profibus interface if present PRO...

Страница 35: ...PE parameter continuous or may be active only at the time of starting INTERLOCK TYPE parameter impulsive START STOP LEVEL The START STOP input acts as only START command active on the level The INTERL...

Страница 36: ...ime in hh mm ss measured from start within which the SETPOINT R1 R2 is not driven even if the threshold value has been reached This parameter is not available if SETPOINT TYPE is set to NORMAL R1 R2 O...

Страница 37: ...600 00 1200 01 4800 9600 31 Flow 6 6 SERIAL F 1 BAUD RATE permits selection of the communication speed selecting this from 600 1200 4800 9600 baud SERIAL MODE permits setting of the type of interface...

Страница 38: ...orderable parts table 6 Ready to operate by SERIAL connections 1 2 3 4 5 6 7 8 9 15 8 13 6 12 5 14 11 7 4 2 1 9 Open Stop Cloed Start Imput Connector Output Connector Controller Start Stop Low Speed...

Страница 39: ...ProfiBus connector Configuration Before power on the controller set the ProfiBus address by the 2 rotary switches Up to 126 different addresses from 0 to 125 can be selected The address value sets by...

Страница 40: ...n state WAIT_PRM or SRC_ADDR Message mapping The communication is based on the continuous data exchange capability of the ProfiBus standard where the master sends continuously the entire slave configu...

Страница 41: ...er Buffer Offset 19 B0 0 0 infinite the valve will never open 126 11 2 0 2sec bit 0 30000 Vent Valve Open Time Set how long the vent will remain open then both purge and vent will be closed Valid only...

Страница 42: ...2 Reserved 19 1 B3 4 Start Stop Interlock Mode Change the logic of Start Stop and Interlock inputs Ask to Varian personnel for this option 151 B0 Vent Purge Auto Man 0 Auto the controller drives purge...

Страница 43: ...ion WIN Serial B0 2 Setpoint R1 Type 0 frequency 1 current 2 time 3 normal operation 4 pressure 101 B3 5 Setpoint R2 Type 0 frequency 1 current 2 time 3 normal operation 4 pressure 171 21 1 B6 7 Analo...

Страница 44: ...the pump has been stopped for at least3 months 0 On 1 Off 100 B4 Vent Valve Command 0 Close 1 Open Valid only if in Par Buff Offset 19 B0 1 122 B3 Purge Valve Command 0 Close 1 Open Valid only if Par...

Страница 45: ...I O connector missing 2 Ramp the pump is increasing the speed 3 Auto tuning the speed has been reduced because the gas load is higher than the Power Limit Applied 4 Braking the pump is reducing the sp...

Страница 46: ...Power Motor power adsorption 201 11 2 W 0 200 Power Limit Applied Maximum allowable power 155 13 2 Hz 0 1500 Driving Frequency Read the rotational speed of the pump 203 15 2 Hz 0 1500 Target Frequenc...

Страница 47: ...al diagnostic service is required and one byte in the diagnostic buffer signals the type of failure User parameter mismatch if the user sets one or more parameters out of their allowed range a ProfiBu...

Страница 48: ...switched off and the display will be as shown in the following figure T O O H I G H L O A D Verify that systems has no leaks than press STOP RESET pushbutton twice to start the pump Check that pump ro...

Страница 49: ...to start the pump ORDERABLE PARTS DESCRIPTION PART NUMBER Turbo V 81 AG 969 8988 Turbo V 81 AG with RS 232 485 option 969 8989 Turbo V 81 AG with Profibus option 969 8990 Turbo V 81 AG Air Cooling Ki...