99 87-900-986-01 (
C
)
Parameter Buffer
Offset
Byte
Size
Byte
Unit
Res
Range
Description
WIN Serial
0
1
-
0-99
Gauge gas load type
The value is divided by 10 (entering 12 the controller will
apply 1.2)
Enter here the correction factor for gas (necessary only if
gauge is connected).
1
2
mA/Hz/sec
0-10000
Setpoint 1 value
Enter here the threshold value for the R1 output, valid
only if B
0
-B
2
of Parameter Buffer Offset-20 is equal to
0,1,2 or 3.
102
3
2
Sec
0-30000
Setpoint 1 Mask Time
Set here the time, measured from the start pump event,
while the R1 output is masked (not activated)
103
5
2
mA/Hz/sec
0-10000
Setpoint 2 value
Enter here the threshold value for the R2 output, valid
only if B
3
-B
5
of Parameter Buffer Offset-20 is equal to
0,1,2 or 3.
172
7
2
Sec
0-30000
Setpoint 2 Mask Time
Set here the time, measured from the start pump event,
while the R2 output is masked (not activated)
173
9
2
0.2sec/bit
0-30000
Vent Valve Opening Delay
Set the delay time between the stop command and vent
opening. Valid only in “purge/vent auto mode”
(Parameter Buffer, Offset 19, B
0
= 0). 0 = infinite (the
valve will never open).
126
11
2
0.2sec/bit
0-30000
Vent Valve Open Time
Set how long the vent will remain open, then both purge
and vent will be closed. Valid only in “purge/vent auto
mode” (Parameter Buffer, Offset 19, B
0
= 0). 0 = infinite
(the valve remain open till next start command).
147
13
1
%
1-99
Setpoint 1 Hysteresis
Set the Setpoint1 hysteresis around the nominal value
105
14
1
%
1-99
Setpoint 2 Hysteresis
Set the Setpoint2 hysteresis around the nominal value
175
15
1
mBar/Torr/
Pa
0-99
Setpoint 1 Pressure Value Mantissa
Set the mantissa of the exponential value of the R1
setpoint (format X.XE-XX) when it is used for pressure
monitoring (Parameter Buffer, Offset 20, B
0-2
= 4). The
value set here is divided by 10, that means 11 = 1.1
162
16
1
-10 - +10
Setpoint 1 Pressure Value Exponent
Set the exponent of the exponential value of the R1
setpoint (format X.XE-XX) when it is used for pressure
monitoring (Parameter Buffer, Offset 20, B
0-2
= 4).
162
17
1
mBar/Torr/
Pa
0-99
Setpoint 2 Pressure Value Mantissa
Set the mantissa of the exponential value of the R2
setpoint (format X.XE-XX) when it is used for pressure
monitoring (Parameter Buffer, Offset 20, B
3-5
= 4). The
value set here is divided by 10, that means 11 = 1.1
176