TECHNICAL INFORMATION
80 87-900-986-01 (
C
)
CHARACTER
ASCII
CONVERTED ASCII
FOR CRC CALCULATION
"A"
CRC
41h
-
-
BFh
"B"
CRC
42h
-
-
BEh
"C"
CRC
43h
-
-
BDh
"D"
CRC
44h
-
-
BCh
"E"
CRC
45h
-
-
BBh
"F"
CRC
46h
-
-
BAh
"G"
CRC
47h
-
-
B9h
"H"
48h
-
"I "
CRC
49h
-
-
B7h
"J"
CRC
4Ah
-
-
B6h
"K"
CRC
4Bh
-
-
B5h
"ACK"
CRC
06h
-
-
FAh
"NACK"
CRC
15h
-
-
EBh
Answer
= after a request from the host, the
Turbo-V 81-AG controller will answer in one of the following
ways:
•
ACK
•
NACK
•
Message
When the Request is "A, "B”, "C", "D", "F", the Turbo-V 81-
AG controller will Answer the ACK or NACK.
When the Request is "E, the
Answer
will contain the
complete set of the following parameters:
BYTES
MEANING
1
MSB LSB
- - - - X X X X
0 = STOP
1 = WAITING INTERLOCK
2 = STARTING
3 = NORMAL OPERATION
4, 5 = HIGH LOAD
6 = FAILURE
7 = APPROACHING LOW SPEED
2
÷
5
Cycle Time
6
÷
9
Pump Life
10
÷
11
Pump Temperature
12
Current*
13
Voltage*
14
÷
17
Frequency
18
÷
19
Cycle #
20
R1 Status
21
R2 Status
22
CRC
*The values for current and voltage are given as numbers,
scaled from 0 to 255, where 0 corresponds to 0 V and 255
to the full scale voltage (130 V) or current (2.5 A).
When the
Request
is "G", the
Answer
will contain a string
of 11 characters with the following parameters:
BYTES
MEANING
1-2
Pump cycles number
(integer coded in 2 bytes)
3
Speed threshold
4-7
Run up time in seconds
(long coded in 4 bytes)
8
Deat time
(0 = NO 1 = YES)
9
Reserved
10
Soft Start mode
(0 = NO 1=YES)
11
CRC
When the configuration parameters have to be changed,
send a
Request
string a string with 9 characters of the
following type:
"H" + .... data .... + CRC