User's Guide _______________________________________________________________________
82 ___________________________________________________________________ M210470EN-B
parameters match, the device
responds with its address. The
settings can now be changed using
aXU! (SDI-12) or aXU<cr><lf>
(ASCII/NMEA) commands. A
software/hardware reset is needed
to validate the changes.
Connection works but data
messages not available.
Wrong device address in a SDI-
12 command or a mistyped
SDI-12 command (in SDI-12 a
mistyped command is in no
way responded).
A command mistyped in
ASCII/NMEA mode while
error messaging/text messages
is disabled (aSU,S=N).
Request the device address with ?!
command and then retype the
command now with the correct
address. Check the data query
commands from Chapter 7,
Getting
data messages
.
Enable the error messaging using
the WXT Configuration Tool or
any terminal by setting aSU,S=Y,
then try the command again.
Data messages are not in
expected format.
The communication protocol
may not be the one you want.
Check the communication protocol
of the device by using the WXT
Configuration Tool or any terminal
with command aXU,M! (SDI-12)
aXU,M<cr><lf> (ASCII/NMEA)
and change it if needed. See
Chapter 6,
Changing the
communication settings
.
Some parameters are
missing from the data
messages.
The formatting of the data
messages is not what you
expect.
Format the data messages of the
concern by using the WXT
Configuration Tool or any terminal
program. See Chapter 8,
Sensor
and data message settings.
An error message as a
response to a command.
See the Section Error
messaging/text messages.
See the section Error
messaging/text messages.