Chapter 7 ___________________________________________________ Getting the data messages
VAISALA________________________________________________________________________ 47
where
<cr><lf>
=
Response terminator
NOTE
For changing the message parameters, units and other settings, see
Chapter 8,
Sensor and data message settings.
NOTE
When the measurement takes less than one second, the response part
2 is not sent. In the WXT510 this is the case in the precipitation
measurement
aM3
.
NOTE
The maximum number of parameters that can be measured with
aD
command is 9. If more parameters are to be measured, command
Start concurrent measurement
aC
should be used, see the sections
below.
Start measurement command with
CRC (aMC)
Command format:
aMCx!
The function of this command, the definition of x, the limitations and
the response structure are the same as for the Start measurement
command
aM
.
In order to request the measured data, Send data command
aD
should
be used, see the following sections.
Start concurrent measurement (aC)
This command is used when there are several devices on the same bus
and simultaneous measurements are needed from the devices. Or if
more than 9 measurement parameters are requested from the single
device.
The measured data is not sent automatically and it should be requested
with separate Send data command
aD!
See examples on page 50.
Command format:
aCx!