Chapter 8 _____________________________________________ Sensor and data message settings
VAISALA________________________________________________________________________ 75
In SDI-12 mode a separate enquiry (0XRU!) must be given to check
the data content.
Supervisor message
Checking the settings
With this command you can check the current supervisor
settings.
Command format in ASCII and NMEA 0183:
aSU
<cr><lf>
Command format in SDI-12:
aXSU
!
where
a
=
Device address (default = 0)
SU
=
Supervisor settings command in ASCII and NMEA
0183
XSU
=
Supervisor settings command in SDI-12
<cr><lf>
=
Command terminator in ASCII and NMEA 0183
!
=
Command terminator in SDI-12
The response in ASCII and NMEA 0183:
aSU,R=[R],I=[I],S=[S],H=[Y]<cr><lf>
The response in SDI-12:
aXSU,R=[R],I=[I],S=[S],H=[Y]<cr><lf>
Setting fields
[R]
=
Parameter selection
: This field consists of 16 bits defining
the supervisor parameters included in the data messages.
The bit value 0 disables and the bit value 1 enables the
parameter.
The parameter order is shown in the following table: