![Vaisala WXT510 Скачать руководство пользователя страница 43](http://html1.mh-extra.com/html/vaisala/wxt510/wxt510_user-manual_849366043.webp)
Chapter 7 ___________________________________________________ Getting the data messages
VAISALA________________________________________________________________________ 43
Example:
0R1,Dm=027D,Sm=0.1M<cr><lf>
0R2,Ta=74.6F,Ua=14.7P,Pa=1012.9H<cr><lf>
0R3,Rc=0.10M,Rd=2380s,Ri=0.0M,Hc=0.0M,Hd=0s,Hi=0.0M<cr><
lf>
0R5,Th=76.1F,Vh=11.5N,Vs=11.5V,Vr=3.510V<cr><lf>
Example (with CRC):
0r1,Sn=0.1M,Sm=0.1M,Sx=0.1MGOG
0r2,Ta=22.7C,Ua=55.5P,Pa=1004.7H@Fn
0r3,Rc=0.00M,Rd=0s,Ri=0.0MIlm
0r5,Th=25.0C,Vh=10.6#,Vs=10.8V,Vr=3.369VO]T
NOTE
Stop the automatic output by changing the communication protocol to
polled mode (aXU,M=P).
SDI-12 protocol
For changing the message parameters, units and other settings, see
Chapter 8
Sensor and data message settings.
Acknowledge active command (a)
This command is used to ensure that a device is responding to a data
recorder or another SDI-12 device. It asks a device to acknowledge its
presence on the SDI-12 bus.
Command format:
a!
where
a
=
Device address (default = 0)
!
=
Command terminator
The response:
a
<cr><lf>
where
a
=
Device address (default = 0)
<cr><lf>
=
Response terminator
Example:
0!
0
<cr><lf>