Chapter 6 ______________________________________________________ Communication settings
VAISALA________________________________________________________________________ 31
NOTE
Reset the transmitter to validate the changes of communication
parameters by disconnecting the service cable or using the Reset
command, page 32.
Example (ASCII and NMEA 0183, device
address 0):
Changing the device address from 0 to 1:
0XU,A=1<cr><lf>
1XU,A=1<cr><lf>
Checking the changed settings:
1XU<cr><lf>
1XU,A=1,M=P,C=2,B=19200,D=8,P=N,S=1,L=25,V=1.00<cr><lf>
Example (ASCII, device
address 0):
Changing RS-232 serial interface with ASCII, polled communication
protocol and baud settings 19200, 8, N, 1 to RS-485 serial interface
with ASCII, automatic protocol and baud settings 9600, 8, N, 1.
Checking the actual settings:
0XU<cr><lf>
0XU,A=0,M=P,C=2,B=19200,D=8,P=N,S=1,L=25,V=1.00<cr><lf>
NOTE
You can change several parameters in the same command as long as
the command length does not exceed 32 characters.
You do not have to type those setting fields that are not to be
changed.
Changing several settings with one command:
0XU,M=A,C=3,B=9600<cr><lf>
0XU,M=A,C=3,B=9600
<cr><lf>
Checking the changed settings:
0XU
<cr><lf>
0XU,A=0,M=A,C=3,B=9600,D=8,P=N,S=1,L=25,V=1.00
<cr><l
f>