One-Point Dewpoint Adjustment using
Serial Line
1.
Let the sensor stabilize for at least 5 hours.
2. Use
the
PURGE
command to start the manual purge. See section
3.
Wait at least 1 hour. Then press adjustment button on the
motherboard inside the transmitter to enable adjustments.
4. Use
the
IKØ
command to perform the adjustment.
IKØ
<cr>
Example:
>
ik0
Tf : -19.74 'C Ref ?
-60
Wait for AutoCal data...OK
>
5.
Press the adjustment button on the motherboard inside the
transmitter to disable adjustments.
NOTE
Several AutoCals can be necessary after this adjustment has been
performed until the transmitter reaches full accuracy.