A sensor purge should be performed always before calibration (see the
calibration instructions) or when there is a reason to believe that a sensor
has become exposed to an interfering chemical.
Starting and Configuring Sensor Purge
Using Display/Keypad (Optional)
1. Open
the
Main Menu
by pressing any of the
▼▲◄►
arrow
buttons.
2. Select
►
Measuring
, press
►
button
.
3. Select
►
Purge
, press
►
button.
1103-067
Figure 73
Sensor Purge Settings
- Start the sensor purge manually by selecting
Start purge now
and pressing START.
- Select
Purge on power-up
by using the arrow buttons. Press
On/Off to turn the start-up purge on/off.
- Set the automatic purge interval by selecting
Interval: ...
, press
SET.
Set the purge interval and the unit (minutes/hours) by
using the arrow buttons. The interval must be 10 min … 48 h.
Press
OK.
- If you have the data logger module installed, you can also
specify the
Date
and
Time
when the purge is performed next.
4. Press
EXIT
to return to the basic display.
1103-066
Figure 74
Performing Sensor Purge