Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 99
Getting the Measurement Message from
Serial Line
Starting Continuous Outputting
Enter the
R
command to start the continuous output of measurements.
R
<cr>
Example:
>
r
Tdf=-20.6 'C H2O= 958 ppmV x= 0.6 g/kg
>
r
Tdf=-20.7 'C H2O= 958 ppmV T= 23.8 'C RH= 3.3 %RH
If a value is too long to fit to the allocated space in the output, or if there
is an error in outputting the quantity, the value is displayed with stars ‘*’.
You can change the format of the output with the following commands:
- Outputting interval can be changed with the
INTV
command.
- Output message format can be changed with the
FORM
command.
Stopping Continuous Outputting
Use the
S
command to end the RUN mode. After this command all other
commands can be used. You can also press the Esc button or reset the
transmitter to stop the outputting.
S
<cr>
See command
SMODE
to change the default (power-up) operation
mode.
Outputting Reading Once
Use the
SEND
command to output the reading once in STOP mode. The
output format depends on which parameters the transmitter can output.
SEND
<cr>
Examples:
>
send
Tdf= -5.8 'C H2O= 3715 ppmV x= 2.3 g/kg