Two-Point Relative Humidity Adjustment
using Serial Line
Perform purge before adjustment. Use the
PURGE
command. See
Then press adjustment button on the motherboard inside the transmitter
to enable adjustments
Perform adjustment with the
FCRH
command. Wait at least 1 hour
before adjusting each point.
FCRH
<cr>
Example:
>
fcrh
RH : 3.90 Ref1 ?
0
Press any key when ready ...
RH : 3.90 Ref2 ?
15
OK
>
One-Point Dewpoint Adjustment using
Display/Keypad
Before proceeding with the T
d/f
adjustment you must perform the RH
adjustment. For DMT340 DRYCAP® M-sensor, the reference
temperature of the dewpoint should be -57 ˚C ... -67 ˚C (-70.6 ˚F ... -88.6
˚F).
Use preferably the display/keypad to perform one-point dewpoint T
d/f
adjustment. To adjust the dewpoint, follow the instructions below:
1. Press
the
ADJ
button on the motherboard to open the
ADJUSTMENT MENU
.
2. Press
the
►
arrow button to select
Adjust T
d
measurement
.
3.
Select by pressing
1-point T
d
adjustment
. Press
START
to start
adjusting.
4.
Leave the sensor to stabilize for at least 5 hours. Follow the
instructions on the display.