page 100 of 121
Tools required (metric):
- Diaphragm key width 46 mm
- Open end wrench width 14 / 17 mm
- 4 / 5 mm wide Allen key
- 2.5 mm wide slotted screwdriver
- Phillips screwdriver size 2
- Torx driver T10
- Flat pliers
Service kit MD 1C
(diaphragms, valves, diaphragm key) ..............................
696828
Ensure that maintenance is done only by suitably
trained and supervised technicians.
The valves and diaphragms are wear parts. If the rated
ultimate vacuum is no longer achieved or in case of in-
creased noise level, the pump interior, the diaphragms
and the valves must be cleaned and the diaphragms and
valves must be checked for cracks or other damage.
All bearings are encapsulated and are filled with long-life
lubricant. Under normal operating conditions, the drive
system is maintenance free.
In demanding circumstances, it may be efficient to check
and clean the pump heads on a regular basis. In normal
use, the lifetime of the diaphragms and valves is typically
15,000 operating hours.
- Prevent internal condensation, transfer of liquids or
dust. The diaphragms and valves will be damaged if
liquid is pumped in significant amount.
- Carry out maintenance frequently if the pump is ex-
posed to corrosive media or in case of deposits.
- Regular maintenance will improve the lifetime of the
pump and also protect both users and the environment.
NOTICE
Содержание PC 3001 TE VARIO PRO
Страница 32: ...page 32 of 121 PC 3001 VARIOpro 1 5 2 9 7 11 12 4 3 10 6 8 14 17...
Страница 33: ...page 33 of 121 PC 3001 TE VARIOpro 5 2 10 1 7 13 8 6 14 17 4 3...
Страница 34: ...page 34 of 121 PC 3001 VARIOpro IK 1 5 2 9 7 4 3 10 8 11 12 15 17 11 12 6...
Страница 36: ...page 36 of 121 PC 3001 VARIOpro without exhaust waste vapor condenser 1 5 2 7 4 3 10 8 14 6 with silencer...