page 48 of 121
Clean the pump if necessary to avoid an increase of the
pump’s operating temperature.
Operation with silencer (optional) at the outlet: Operating
the pump at a high inlet pressure or pumping dusty gases
for a long time may cause clogging of the silencer. Check
the silencer regularly and replace if necessary.
A temperature sensor at the circuit board protects the mo-
tor: Current limitation in the event the temperature at the
circuit board raises above 158°F (70°C). At temperatures
above 185°F (85°C) the pump is switched off. In the event
of a motor blockage (after 10 start-up attempts) the pump
is switched off.
Note
: Only manual reset is possible. Disconnect the pump
from the power source. Identify and eliminate the cause of
failure.
Check ventilation slots regularly for dust/dirt. Clean if nec-
essary. Avoid overheating (e.g., due to hot process gases).
A warm up period (approximately 15 min.) is required to
ensure that the rated ultimate vacuum and pumping speed
are attained.
Important notes regarding the use of gas bal-
last
Gas ballast is a continuous purge to keep the pump’s in-
terior as clean as possible and to reduce the possibility of
condensation inside the pump.
➨
Air and pumped media might react inside the pump or
at the outlet of the pump and form hazardous or explo-
sive mixtures, when you use air rather than inert gas
for the gas ballast. This constitutes a risk of significant
damage to equipment and/or facilities, a risk of per-
sonal injury or even loss of life.
Содержание PC 3001 TE VARIO PRO
Страница 32: ...page 32 of 121 PC 3001 VARIOpro 1 5 2 9 7 11 12 4 3 10 6 8 14 17...
Страница 33: ...page 33 of 121 PC 3001 TE VARIOpro 5 2 10 1 7 13 8 6 14 17 4 3...
Страница 34: ...page 34 of 121 PC 3001 VARIOpro IK 1 5 2 9 7 4 3 10 8 11 12 15 17 11 12 6...
Страница 36: ...page 36 of 121 PC 3001 VARIOpro without exhaust waste vapor condenser 1 5 2 7 4 3 10 8 14 6 with silencer...