- Condensate and deposits will affect the mea-
surement results. Clean the pressure trans-
ducer, if necessary. See section „Cleaning
the pressure transducer“, pg. 83.
Connecting components at the controller:
At the rear side of the controller are connec-
tions for e.g., an optional coolant valve and/or
an external pressure transducer and/or an ex-
ternal venting valve, as well as a venting con-
nection and the serial interface RS 232C.
- Pull the controller out of the housing. Be careful not to
disconnect the measuring connection (PTFE tube)!
- Insert the cable of the component to be connected un-
derneath the housing and plug into the controller. Do
not apply off-axis forces when assembling or removing
plug connections! Observe correct orientation of the
plug.
- Insert controller into the housing.
Separator at inlet and exhaust waste vapor condens-
er:
Assembling the hose nozzle with union nut:
➨
Take the hose nozzle with attached compression fer-
rule and union nut out of the catchpot and put onto
inlet connection.
➨
Tighten the union nut by hand until you can feel the
stop. Then tighten an additional 1/4 rotation with an
open-ended wrench (size 17mm) for final installation.
measurement
connection
at CVC 3000
Содержание PC 3001 TE VARIO PRO
Страница 32: ...page 32 of 121 PC 3001 VARIOpro 1 5 2 9 7 11 12 4 3 10 6 8 14 17...
Страница 33: ...page 33 of 121 PC 3001 TE VARIOpro 5 2 10 1 7 13 8 6 14 17 4 3...
Страница 34: ...page 34 of 121 PC 3001 VARIOpro IK 1 5 2 9 7 4 3 10 8 11 12 15 17 11 12 6...
Страница 36: ...page 36 of 121 PC 3001 VARIOpro without exhaust waste vapor condenser 1 5 2 7 4 3 10 8 14 6 with silencer...