page 18 of 121
mospheres
Equipment marking (ATEX)“, pg. 25)
.
➨
The pumps are
not suitable
to pump any of the sub-
stances listed below.
Do not pump:
-
unstable substances
- substances which react explosively under
impact
(mechanical stress) without air
- substances which react explosively when being ex-
posed to
elevated temperatures
without air,
- substances subject to
auto-ignition
,
- substances which are
inflammable without air
-
explosive substances
.
➨
The pumps are
not approved
for operation below
ground.
Do not operate
the pump
below ground
.
•
Do not pump
substances which may form
depos-
its
inside the pump. The pumps are not suitable for
pumping substances which may form deposits inside
the pump. Deposits and condensate in the pump may
lead to increased temperatures even to the point of
exceeding the maximum permitted temperatures.
• Check the inlet and outlet of the pump, if there is a
danger of forming
deposits
inside the pump, e.g., in
the pump chambers (the pump chamber is the part
between diaphragm and head cover. See section „Re-
placing diaphragms and valves“, pg. 99). Inspect the
pump chambers regularly and clean if necessary.
•
Consider interactions and chemical reactions of
the pumped media.
Ensure that the materials of the
pump’s wetted parts are compatible with the pumped
substances, see section „Technical data“, pg. 27.
When changing the substances pumped, we recom-
mend purging the pump with air or inert gas prior to
changing the pumped media. Purging the pump will
pump out residues and it will reduce the possibility of
Содержание PC 3001 TE VARIO PRO
Страница 32: ...page 32 of 121 PC 3001 VARIOpro 1 5 2 9 7 11 12 4 3 10 6 8 14 17...
Страница 33: ...page 33 of 121 PC 3001 TE VARIOpro 5 2 10 1 7 13 8 6 14 17 4 3...
Страница 34: ...page 34 of 121 PC 3001 VARIOpro IK 1 5 2 9 7 4 3 10 8 11 12 15 17 11 12 6...
Страница 36: ...page 36 of 121 PC 3001 VARIOpro without exhaust waste vapor condenser 1 5 2 7 4 3 10 8 14 6 with silencer...